BioSafety Cabinet Class II A2 & B2 

BioSafety Cabinet

Kapsam Class II Type A2 and Type B2 microbiological safety (biosafety) cabinets provide superior protection for products, personnel, and the environment, enabling safe work with toxic chemicals and high pathogens.

In both models, air taken from the environment passes through a HEPA filter and then through a laminator to reach the working area, ensuring product protection with filtered laminar airflow. The air barrier at the front of the device guarantees personnel protection.

In the Class II Type B2 model, all air passes through the HEPA filter at high speed and is 100% exhausted outside.In the Class II Type A2 model, 30% of the air that passes through the glass opening and HEPA filter is exhausted to the external environment through the HEPA filter, while 70% is recirculated back into the work area after passing through the HEPA filter again.

Body & Design Features

Microprocessor
Control Unit

Touch Control Panel Features

Optional Features

Minimize Cross-Contamination
with Waste Sealing System

Kapsam Biosafety Cabinets are optionally integrated with automated waste sealing system, enabling convenient waste disposal within the safe workspace. Integrated with cabinet automation, it enables monitoring foil length and receiving alerts before depletion of foil cassette ensuring the uninterrupted operations and a seamless and safe workflow.

Mobile Sealing System

for more information

visit product web page

Technical

Information

BioSafety Cabinet Class II A2

Model:

KS-BK2A2-90

KS-BK2A2-120

KS-BK2A2-150

KS-BK2A2-180

DIMENSIONS & WEIGHT

Int. Dimension (w x d x h):

920x630x680

1220x630x680

1520x630x680

1820x630x680

Ext. Dimension (w x d x h):

1050x820x1560

1350x820x1560

1650x820x1560

1950x820x1560

Carrying Stand (w x d x h):

1050x780x710

1350x780x710

1650x780x710

1950x780x710

Workbench Height (mm):

770

770

770

770

Cabinet Weight:

300 kg

325 kg

350 kg

375 kg

Carrying Stand Weight:

55 kg

60 kg

65 kg

70 kg

Total Weight:

355 kg

385 kg

415 kg

445 kg

HEPA FILTERS & AIRFLOW

Inflow Airflow Speed:

≥ 0,40 m/s

Downflow Airflow Speed:

0,25 - 0,50 m/s

Pre Filter (EN 1822):

EU4

Main & Exhaust Filter (EN 1822):

H14 HEPA, 0.3 µm particle 99.995%< filtration efficiency

Cleanliness Class (EN ISO 14644-3):

< ISO 5

Cleanliness Class (US FED 209E):

Class 100

ELECTRICAL PARAMETERS

Fan Filter Unit (FFU)

250 W

250 W

250 W

500 W

Energy Consumption (230 VAC - 50 Hz)

Lightening

12 W

16 W

20 W

24 W

UV Lamp

15 W

15 W

15 W

15 W

Electrical Socket

2 x (1150 W / 5 A)

Total Power / Current:

3427W / 14,9A

3430W / 14,9A

3434W / 14,9A

3639W / 15,8A

Power Consumption under Normal Operating Conditions (Lightening & Fan Motor):

750 W

780 W

750 W

925 W

ERGONOMY & COMFORT

Windshield Opening
(Standard (Pre-set) / Maximum):

200 ± 10 mm /
440 ± 10 mm

Noise Level & Lighting Intensity:

65 dB & 800 lux

MATERIAL

Workbench (Standard / Optional):

304 Stainless Steel / 316 Stainless Steel

Exterior Material:

Antibacterial electrostatic paint over steel material.

Front Windshield Thickness:

6 mm (Tempered glass)

Technical

Information

BioSafety Cabinet Class II B2

Model:

KS-BK2B2-90

KS-BK2B2-120

KS-BK2B2-150

KS-BK2B2-180

DIMENSIONS & WEIGHT

Int. Dimension (w x d x h):

920x630x680

1220x630x680

1520x630x680

1820x630x680

Ext. Dimension (w x d x h):

1050x820x1670

1350x820x1670

1650x820x1670

1950x820x1670

Carrying Stand (w x d x h):

1050x780x710

1350x780x710

1650x780x710

1950x780x710

Workbench Height (mm):

770

770

770

770

Cabinet Weight:

300 kg

325 kg

350 kg

375 kg

Carrying Stand Weight:

55 kg

60 kg

65 kg

70 kg

Total Weight:

355 kg

385 kg

415 kg

445 kg

HEPA FILTERS & AIRFLOW

Inflow Airflow Speed:

≥ 0,40 m/s

Downflow Airflow Speed:

0,25 - 0,50 m/s

Pre Filter (EN 1822):

EU4

Main & Exhaust Filter (EN 1822):

H14 HEPA, 0.3 µm particle 99.995%< filtration efficiency

Cleanliness Class (EN ISO 14644-3):

< ISO 5

Cleanliness Class (US FED 209E):

Class 100

Exhaust Fan Motor Capacity:

920 m³/h

1075 m³/h

1340 m³/h

1603 m³/h

ELECTRICAL PARAMETERS

Fan Filter Unit (FFU)

208 W

208 W

416 W

416 W

Energy Consumption (230 VAC - 50 Hz)

Lightening

12 W

16 W

20 W

24 W

UV Lamp

15 W

15 W

15 W

15 W

Electrical Socket

2 x (1150 W / 5 A)

Total Power / Current:

3265 W / 14,2 A

3269 W / 14,2 A

3481 W / 15,1 A

3485 W / 15,1 A

Power Consumption under Normal Operating Conditions (Lightening & Fan Motor):

550 W

580 W

750 W

900 W

ERGONOMY & COMFORT

Windshield Opening
(Standard (Pre-set) / Maximum):

200 ± 10 mm /
440 ± 10 mm

Noise Level & Lighting Intensity:

65 dB & 800 lux

MATERIAL

Workbench (Standard / Optional):

304 Stainless Steel / 316 Stainless Steel

Exterior Material:

Antibacterial electrostatic paint over steel material.

Front Windshield Thickness:

6 mm (Tempered glass)